Title
Design of a wide range kVp-meter
Author
Tümer, Murat ; Ülgen, Yekta
Author_Institution
Biyomed. Muhendisligi Enstitusu, Bogazici Univ., Istanbul, Turkey
fYear
2010
fDate
21-24 April 2010
Firstpage
1
Lastpage
4
Abstract
The tube voltage is one of the major factors affecting the image quality in X-ray imaging. It is in kilovolts range and usually called kVp, since it is measured as the peak value during X-ray production. The kVp determines the penetration and absorption characteristics of the X-ray photons in the tissue and contributes directly to image quality. The accuracy and precision of the kVp should be measured annually to be 5%. The most accurate and precise method would be measuring the tube voltage directly with a voltmeter. However, the cost of an equipment for measuring such a high voltage and the danger for the technician makes non-invasive kVp-meters the best alternative. In this work, a previsously designed and prototyped kVp-meter is redesigned and its accuracy and precision at higher voltages improved to below 1% throughout the measurement range. The design makes use of the dependency of the linear attenuation coefficient of metals on the energy of the incident photons, which are in turn related to kVp.
Keywords
X-ray tubes; biological tissues; diagnostic radiography; voltage measurement; voltmeters; X-ray imaging; X-ray photon absorption characteristic; X-ray photon penetration characteristic; X-ray production; image quality; linear attenuation coefficient; metals; tissue; tube voltage; wide range kVp-meter; Attenuation; Costs; Electromagnetic wave absorption; Image quality; Production; Prototypes; Voltage measurement; Voltmeters; X-ray imaging;
fLanguage
English
Publisher
ieee
Conference_Titel
Biomedical Engineering Meeting (BIYOMUT), 2010 15th National
Conference_Location
Antalya
Print_ISBN
978-1-4244-6380-0
Type
conf
DOI
10.1109/BIYOMUT.2010.5479751
Filename
5479751
Link To Document