• DocumentCode
    2586145
  • Title

    Characterisation of a coding/reconstruction algorithm on MRI simulated noisy data

  • Author

    Placidi, Giuseppe ; Franchi, Danilo ; Maurizi, Alfredo

  • Author_Institution
    Dept. of Health Sci., Univ. of LAquila, L´´ Aquila, Italy
  • fYear
    2009
  • fDate
    29-30 May 2009
  • Firstpage
    156
  • Lastpage
    159
  • Abstract
    In Magnetic Resonance Imaging (MRI), the presence of static magnetic field in-homogeneities introduces distortions and artifacts in the images. A coding/reconstruction algorithm for MRI, based on the assignment of different time varying frequencies (accelerations) to different spatial positions, has been tested on simulated noisy data. Results, demonstrating the method is not conditioned neither by the presence of static magnetic field in-homogeneity nor by the presence of noise, are reported. In particular, the results demonstrated that noise was not incremented and the signal was unaffected by blurring or distortions.
  • Keywords
    biomedical MRI; image coding; image reconstruction; medical image processing; MRI simulated noisy data; image coding algorithm; image reconstruction algorithm; magnetic resonance imaging; static magnetic field in-homogeneities; time varying frequency; Acceleration; Biomedical imaging; Decoding; Frequency; Image coding; Magnetic field measurement; Magnetic fields; Magnetic noise; Magnetic resonance imaging; Reconstruction algorithms; MRI; acquisition sequences; coding; decoding; frequency temporal variation; modulated gradients;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Medical Measurements and Applications, 2009. MeMeA 2009. IEEE International Workshop on
  • Conference_Location
    Cetraro
  • Print_ISBN
    978-1-4244-3598-2
  • Electronic_ISBN
    978-1-4244-3599-9
  • Type

    conf

  • DOI
    10.1109/MEMEA.2009.5167975
  • Filename
    5167975