DocumentCode
2589505
Title
Two different atomic oscillators in the same environment: a new technique for compensating perturbations
Author
Seidel, D.J. ; Yei, Wo ; Maleki, L.
Author_Institution
Jet Propulsion Lab., California Inst. of Technol., Pasadena, CA, USA
fYear
1996
fDate
5-7 Jun 1996
Firstpage
1223
Lastpage
1224
Abstract
Presents a new technique for compensating frequency drift in a trapped ion frequency standard by simultaneously holding 171Yb+ and 199Hg+. Environmental perturbations affect the hyperfine splitting of these two ions very differently. By monitoring the changes in the frequency difference between their hyperfine levels we can compensate for some of the frequency drift due to changes in external perturbations
Keywords
atomic clocks; compensation; frequency measurement; frequency stability; hyperfine structure; measurement standards; mercury (metal); microwave measurement; trapped ions; ytterbium; Hg; Yb; atomic oscillators; external perturbations; frequency difference; frequency drift; hyperfine splitting; perturbation compensation; trapped ion frequency standard; Doppler shift; Electrons; Frequency measurement; Isotopes; Laboratories; Magnetic field measurement; Monitoring; Oscillators; Propulsion; Ytterbium;
fLanguage
English
Publisher
ieee
Conference_Titel
Frequency Control Symposium, 1996. 50th., Proceedings of the 1996 IEEE International.
Conference_Location
Honolulu, HI
Print_ISBN
0-7803-3309-8
Type
conf
DOI
10.1109/FREQ.1996.560317
Filename
560317
Link To Document