• DocumentCode
    2591427
  • Title

    An interface chip for power and bidirectional data telemetry in an implantable cochlear microsystem

  • Author

    Sodagar, Amir M. ; Wise, Kensall D. ; Najafi, Khalil

  • Author_Institution
    Center for Wireless Integrated Microsyst., Univ. of Michigan, Ann Arbor, MI
  • fYear
    2006
  • fDate
    Nov. 29 2006-Dec. 1 2006
  • Firstpage
    1
  • Lastpage
    4
  • Abstract
    A wireless interface chip is described that transfers power, data, and clock from an external system to a cochlear implant and feeds back neural-response and position information. The forward telemetry front-end of the chip is comprised of a multi-output power regulator, wideband FSK data demodulator, and clock recovery circuit. The reverse telemetry back end consists of a Manchester encoder, an on-off keying modulator, and an antenna driver. A data packet checker is also considered to check the format and contents (parity) of the received data.
  • Keywords
    bioMEMS; biomedical electronics; biomedical telemetry; demodulators; driver circuits; ear; encoding; frequency shift keying; neurophysiology; prosthetics; Manchester encoder; antenna driver; bidirectional data telemetry; clock recovery circuit; data packet checker; forward telemetry front-end; implantable cochlear microsystem; multioutput power regulator; neural-response; on-off keying modulator; position information; power data telemetry; reverse telemetry back end; wideband FSK data demodulator; wireless interface chip; Bridge circuits; Clocks; Cochlear implants; Demodulation; Driver circuits; Frequency shift keying; Rectifiers; Regulators; Telemetry; Voltage;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Biomedical Circuits and Systems Conference, 2006. BioCAS 2006. IEEE
  • Conference_Location
    London
  • Print_ISBN
    978-1-4244-0436-0
  • Electronic_ISBN
    978-1-4244-0437-7
  • Type

    conf

  • DOI
    10.1109/BIOCAS.2006.4600293
  • Filename
    4600293