• DocumentCode
    2592941
  • Title

    E-field generation setup for UWB-SP sensor calibration

  • Author

    Yan You-jie ; Liu Xiao-long ; Jiang Ting-yong ; Chen Jin ; Li Peng-hui ; Liu Ying

  • Author_Institution
    Northwest Inst. of Nucl. Technol., Xi´an, China
  • fYear
    2012
  • fDate
    21-24 May 2012
  • Firstpage
    541
  • Lastpage
    544
  • Abstract
    In this paper, a mono-cone setup which contain a ground plane and a cone is manufactured to form a standard UWB-SP field. The rigorous analysis of E-field distributing in the monocone has been provided. A D-dot sensor has been calibrated in the monocone. The results of the calibration experiment indicate that the monocone can form an accurately known E-field under the excited of a peak-to-peak width of 500 ps pulse. The measurement results of equivalent area Aeq is consistent with the calculated results and the simulation results.
  • Keywords
    calibration; electric field measurement; electric sensing devices; electromagnetic pulse; ultra wideband antennas; D-dot sensor; E-field distribution; E-field generation setup; UWB-SP field; calibration; electromagnetic pulse; ground plane; monocone setup; Antenna measurements; Attenuation measurement; Attenuators; Bandwidth; Calibration; Impedance; Plasma measurements;