DocumentCode
2592941
Title
E-field generation setup for UWB-SP sensor calibration
Author
Yan You-jie ; Liu Xiao-long ; Jiang Ting-yong ; Chen Jin ; Li Peng-hui ; Liu Ying
Author_Institution
Northwest Inst. of Nucl. Technol., Xi´an, China
fYear
2012
fDate
21-24 May 2012
Firstpage
541
Lastpage
544
Abstract
In this paper, a mono-cone setup which contain a ground plane and a cone is manufactured to form a standard UWB-SP field. The rigorous analysis of E-field distributing in the monocone has been provided. A D-dot sensor has been calibrated in the monocone. The results of the calibration experiment indicate that the monocone can form an accurately known E-field under the excited of a peak-to-peak width of 500 ps pulse. The measurement results of equivalent area Aeq is consistent with the calculated results and the simulation results.
Keywords
calibration; electric field measurement; electric sensing devices; electromagnetic pulse; ultra wideband antennas; D-dot sensor; E-field distribution; E-field generation setup; UWB-SP field; calibration; electromagnetic pulse; ground plane; monocone setup; Antenna measurements; Attenuation measurement; Attenuators; Bandwidth; Calibration; Impedance; Plasma measurements;
fLanguage
English
Publisher
ieee
Conference_Titel
Electromagnetic Compatibility (APEMC), 2012 Asia-Pacific Symposium on
Conference_Location
Singapore
Print_ISBN
978-1-4577-1557-0
Electronic_ISBN
978-1-4577-1558-7
Type
conf
DOI
10.1109/APEMC.2012.6237822
Filename
6237822
Link To Document