Title :
The effect of cryogenic environment on sensitivity of piezoelectric pressure transducers
Author_Institution :
Nat. Key Lab. of Electron. Test Technol., North Univ. of China, Taiyuan, China
Abstract :
Pressure transducer is generally calibrated in room temperature. But in many applications, pressure transducer may be required to work in a very low temperature down to -40°C. Pressure measurement will be affected by extreme temperature in those cases. For example, those real experiments in cold area, transducer and its measurement system often work at very low temperatures. The paper describes a method of determining sensitivity of a piezoelectric pressure transducer in cryogenic environment. Theoretical analysis and experimental results show that the method is useful and effective. As a result, the environmental factor of pressure transducer in cryogenic environment can be calculated.
Keywords :
low-temperature techniques; piezoelectric transducers; pressure measurement; pressure transducers; cryogenic environment; environmental factor; piezoelectric pressure transducer; pressure measurement; very low temperature applications; Cryogenics; Piezoelectric transducers; component; environmental factor; low temperature condition; piezoelectric pressure transducer;
Conference_Titel :
Instrumentation and Measurement Technology Conference, 2009. I2MTC '09. IEEE
Conference_Location :
Singapore
Print_ISBN :
978-1-4244-3352-0
DOI :
10.1109/IMTC.2009.5168439