• DocumentCode
    2595449
  • Title

    The effect of cryogenic environment on sensitivity of piezoelectric pressure transducers

  • Author

    Zhijie Zhang

  • Author_Institution
    Nat. Key Lab. of Electron. Test Technol., North Univ. of China, Taiyuan, China
  • fYear
    2009
  • fDate
    5-7 May 2009
  • Firstpage
    177
  • Lastpage
    180
  • Abstract
    Pressure transducer is generally calibrated in room temperature. But in many applications, pressure transducer may be required to work in a very low temperature down to -40°C. Pressure measurement will be affected by extreme temperature in those cases. For example, those real experiments in cold area, transducer and its measurement system often work at very low temperatures. The paper describes a method of determining sensitivity of a piezoelectric pressure transducer in cryogenic environment. Theoretical analysis and experimental results show that the method is useful and effective. As a result, the environmental factor of pressure transducer in cryogenic environment can be calculated.
  • Keywords
    low-temperature techniques; piezoelectric transducers; pressure measurement; pressure transducers; cryogenic environment; environmental factor; piezoelectric pressure transducer; pressure measurement; very low temperature applications; Cryogenics; Piezoelectric transducers; component; environmental factor; low temperature condition; piezoelectric pressure transducer;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Instrumentation and Measurement Technology Conference, 2009. I2MTC '09. IEEE
  • Conference_Location
    Singapore
  • ISSN
    1091-5281
  • Print_ISBN
    978-1-4244-3352-0
  • Type

    conf

  • DOI
    10.1109/IMTC.2009.5168439
  • Filename
    5168439