DocumentCode
2595449
Title
The effect of cryogenic environment on sensitivity of piezoelectric pressure transducers
Author
Zhijie Zhang
Author_Institution
Nat. Key Lab. of Electron. Test Technol., North Univ. of China, Taiyuan, China
fYear
2009
fDate
5-7 May 2009
Firstpage
177
Lastpage
180
Abstract
Pressure transducer is generally calibrated in room temperature. But in many applications, pressure transducer may be required to work in a very low temperature down to -40°C. Pressure measurement will be affected by extreme temperature in those cases. For example, those real experiments in cold area, transducer and its measurement system often work at very low temperatures. The paper describes a method of determining sensitivity of a piezoelectric pressure transducer in cryogenic environment. Theoretical analysis and experimental results show that the method is useful and effective. As a result, the environmental factor of pressure transducer in cryogenic environment can be calculated.
Keywords
low-temperature techniques; piezoelectric transducers; pressure measurement; pressure transducers; cryogenic environment; environmental factor; piezoelectric pressure transducer; pressure measurement; very low temperature applications; Cryogenics; Piezoelectric transducers; component; environmental factor; low temperature condition; piezoelectric pressure transducer;
fLanguage
English
Publisher
ieee
Conference_Titel
Instrumentation and Measurement Technology Conference, 2009. I2MTC '09. IEEE
Conference_Location
Singapore
ISSN
1091-5281
Print_ISBN
978-1-4244-3352-0
Type
conf
DOI
10.1109/IMTC.2009.5168439
Filename
5168439
Link To Document