DocumentCode
2596378
Title
Calibration of Rogowski coils at frequencies up to 10 kHz using digital sampling
Author
Djokic, Branislav
Author_Institution
Inst. for Nat. Meas. Stand., Nat. Res. Council of Canada, Ottawa, ON, Canada
fYear
2009
fDate
5-7 May 2009
Firstpage
437
Lastpage
441
Abstract
A high-accuracy system for calibration of Rogowski coils at frequencies up to 10 kHz, based on a reference current transformer, an I/V converter, and a digital sampling system, is described in the paper. The calibration system has been automated and its performance evaluated. At the frequency of 10 kHz, the calibration system is capable of generating currents up to 100 A. At power frequencies, it is capable of generating currents up to 2,000 A. The best uncertainties (k=2) of the calibration system are estimated to be less than 500 μA/A for magnitude and 500 μrad for phase at 10 kHz, and an order of magnitude better at power frequencies.
Keywords
calibration; coils; current transformers; power convertors; signal sampling; I/V converter; Rogowski coils; calibration; digital sampling; frequency 10 kHz; output signal digitization; power frequency; reference current transformer; Calibration; Coils; Current transformers; Digital-to-frequency converters; Frequency conversion; Frequency estimation; Phase estimation; Power generation; Sampling methods; Uncertainty; Rogowski coil; calibration; current transformer; digital sampling;
fLanguage
English
Publisher
ieee
Conference_Titel
Instrumentation and Measurement Technology Conference, 2009. I2MTC '09. IEEE
Conference_Location
Singapore
ISSN
1091-5281
Print_ISBN
978-1-4244-3352-0
Type
conf
DOI
10.1109/IMTC.2009.5168488
Filename
5168488
Link To Document