• DocumentCode
    2596487
  • Title

    Thermal stress on silver conductive adhesive solder joints: Performance evaluation of medical ultrasound array transducer

  • Author

    Catelani, Marcantonio ; Scarano, Valeria L. ; Bertocci, Francesco ; Singuaroli, Roberto ; Palchetti, Paolo ; Grandoni, Andrea

  • Author_Institution
    Dept. of Electron. & Telecommun., Univ. of Florence, Florence, Italy
  • fYear
    2009
  • fDate
    5-7 May 2009
  • Firstpage
    468
  • Lastpage
    471
  • Abstract
    RoHS European Directive, recently introduced, makes a restriction of hazardous substances, e.g. Lead (Pb) and so on, traditionally used in soldering process of electric components and equipments. The paper presents preliminary results concerning new electronic Pb-free solder joints, in terms of both joint performances and electronic properties. A particular application of silver conductive adhesive solder joints between a plate of piezoelectric material and fingers, brass micro-connections between the substrate of piezoelectric material and the printed circuit, is considered. Factors that influence the soldering process like “thickness of the conductive glue” and “temperature of polymerization” are analysed by means of the Design of Experiments technique. The aim is to realise samples for the experimental tests. Measurements of electrical resistance, parameter chosen like failure pointer, and morphologic and micro-chemical analyses were carried out in order to verify the samples behaviour to the thermal stresses.
  • Keywords
    biomedical transducers; biomedical ultrasonics; conductive adhesives; design of experiments; electric resistance measurement; performance evaluation; silver; solders; thermal stresses; ultrasonic transducer arrays; Ag; RoHS European Directive; brass microconnection; conductive glue thickness; design-of-experiments technique; electrical resistance measurement; electronic lead-free solder joint; medical ultrasound array transducer; performance evaluation; piezoelectric material plate; polymerization temperature; silver conductive adhesive solder joint; thermal stress; Biomedical transducers; Conductive adhesives; Lead; Silver; Soldering; Thermal conductivity; Thermal stresses; Ultrasonic imaging; Ultrasonic transducer arrays; Ultrasonic transducers; conductive glue; piezoelectric transducer; reliability characterization; silver conductive adhesive; solder joint; thermal tests;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Instrumentation and Measurement Technology Conference, 2009. I2MTC '09. IEEE
  • Conference_Location
    Singapore
  • ISSN
    1091-5281
  • Print_ISBN
    978-1-4244-3352-0
  • Type

    conf

  • DOI
    10.1109/IMTC.2009.5168494
  • Filename
    5168494