DocumentCode
2609016
Title
Design and fabrication of a transceiver for microwave tomography breast imaging system
Author
Kim, Hyuk-Je ; Lee, Jong-Moon ; Jeon, Soon-Ik ; Choi, Hyung-Do
Author_Institution
Bio Electromagn. Res. Team, Electron. & Telecommun. Res. Inst. (ETRI), Daejeon, South Korea
fYear
2012
fDate
15-17 Oct. 2012
Firstpage
238
Lastpage
239
Abstract
A transceiver subsystem for microwave breast imaging system was designed and fabricated. The subsystem consists of a transmitter, a receiver and a 16-channel switch matrix module. The subsystem is operated in the frequency from 500 MHz to 3 GHz and has 130 dB dynamic range and 110 dB channel isolation.
Keywords
medical image processing; microwave imaging; tomography; transceivers; channel isolation; microwave breast imaging system; microwave tomography breast imaging system; receiver; transceiver design; transceiver fabrication; transceiver subsystem; transmitter; Breast; Microwave antennas; Microwave imaging; Microwave theory and techniques; Switches; Tomography; Transceivers; Transceiver; breast cancer; microwave tomography; switch matrix;