DocumentCode :
2610568
Title :
A new beamforming technique aimed at improving the ultrasound harmonic imaging
Author :
Trucco, Andrea ; Bertora, Franco
Author_Institution :
Dept. of Biophys. & Electron. Eng., Genoa Univ., Genova
fYear :
2004
fDate :
14-14 May 2004
Firstpage :
71
Lastpage :
75
Abstract :
In this paper, a modification to the delays used in reception beamforming is introduced that allows one to sharply reduce the echoes centered at the fundamental frequency and to keep the echoes centered at the second harmonic, while preserving the focusing performances. Such a modification can be easily implemented in echographic equipments. The spectral analysis of the obtained signals and the performance reported in processing real signals increase the confidence in the practical value of the proposed technique
Keywords :
biomedical ultrasonics; echo suppression; medical image processing; spectral analysis; ultrasonic imaging; echographic equipment; reception beamforming; signal processing; spectral analysis; ultrasound harmonic imaging; Acoustic imaging; Array signal processing; Delay; Filtering; Frequency; Matched filters; Power harmonic filters; Signal processing; Signal resolution; Ultrasonic imaging;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Imaging Systems and Techniques, 2004. (IST). 2004 IEEE International Workshop on
Conference_Location :
Stresa
Print_ISBN :
0-7803-8591-8
Type :
conf
DOI :
10.1109/IST.2004.1397285
Filename :
1397285
Link To Document :
بازگشت