• DocumentCode
    2611742
  • Title

    A cone beam CT scanner without moving parts

  • Author

    Schomberg, Hermann ; Luhta, Randy ; Pietig, Rainer

  • Author_Institution
    Philips Research Europe - Hamburg, Germany
  • fYear
    2008
  • fDate
    19-25 Oct. 2008
  • Firstpage
    3920
  • Lastpage
    3922
  • Abstract
    We propose a cone beam CT scanner without moving parts. The proposed scanner has a large number of small, stationary x-ray sources that are aligned along a specially shaped source curve. It also has a stationary detector strip with a large area and a special shaping. The sources are based on carbon nanotube field emitters. The detector is assembled from smaller modules having a flat, rectangular detector area. The special shapings of the source curve and the detector strip make it possible to collect a complete set of cone beam projections of a patient’s volume of interest in a very short time. The proposed scanner appears to be well suited for cardiac CT.
  • Keywords
    Assembly; Carbon nanotubes; Computed tomography; Electron beams; Europe; Optical imaging; Particle beams; Strips; X-ray detection; X-ray detectors;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2008. NSS '08. IEEE
  • Conference_Location
    Dresden, Germany
  • ISSN
    1095-7863
  • Print_ISBN
    978-1-4244-2714-7
  • Electronic_ISBN
    1095-7863
  • Type

    conf