Title
Quantitative accuracy of SPECT reconstructions with a voxelized CT-based partial volume effect correction
Author
Shcherbinin, S. ; Celler, A.
Author_Institution
Medical Imaging Research Group, Department of Radiology, University of British Columbia, Vancouver, Canada
fYear
2008
fDate
19-25 Oct. 2008
Firstpage
4311
Lastpage
4318
Abstract
For many clinical applications, it is important to know not only the total activity in an organ of interest, but also the details regarding this activity distribution. In this context, the term “quantitation” would refer to the recovery of not only the total activity but also its voxelized concentration. In this paper, we analyze the quantitative accuracy of activity distributions in 99mTc SPECT/CT images using data from physical phantom experiments. Our acquisitions followed cardiac (99mTc-MIBI) and oncology (99mTc-MDP) clinical acquisition protocols and were aiming at providing the medical community with information about absolute 99mTc activity distribution in the myocardium (for the diagnosis of a balanced triple vessel disease) and in tumors (for evaluation of their aggressiveness). Our algorithm includes the following steps: (i) Reconstruction of initial images with corrections for attenuation, scattering, and resolution loss; (ii) Creation of a template (numeric phantom) covering not only the investigated region of interest (ROI), but also other active areas. allowing us to simultaneously take into account both spill-in and spill-out effects; (iii) Projection and reconstruction of this template (with the same corrections as for the initial images); (iv) Voxel-by-voxel partial volume effect correction (PVC). After applying all the aforementioned corrections, we recovered total activities inside both heart and tumor models with errors less than ∼6% for 33ml containers and ∼9% for 120ml heart insert. PVC substantially improved not only the accuracy of the recovered total activity (up to 40%), but also its distribution inside the ROI. The current version of this PVC technique makes an assumption that the ratios of activity between different compartments are known. Although this assumption may hold in some clinical situations, for example for tumors in areas with low background activity, a- - t the next stage we plan to test the use of the initial image reconstruction for creation of the template.
Keywords
Biomedical imaging; Computed tomography; Heart; Image analysis; Image reconstruction; Imaging phantoms; Medical diagnostic imaging; Neoplasms; Oncology; Protocols; SPECT; accuracy; correction; image; partial volume effect; phantom; quantitation; reconstruction;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium Conference Record, 2008. NSS '08. IEEE
Conference_Location
Dresden, Germany
ISSN
1095-7863
Print_ISBN
978-1-4244-2714-7
Electronic_ISBN
1095-7863
Type
conf
DOI
10.1109/NSSMIC.2008.4774238
Filename
4774238
Link To Document