• DocumentCode
    2614276
  • Title

    Accuracy of channelized hotelling observer performance measures estimated from repeated CT scans

  • Author

    Wunderlich, Adam ; Noo, Frédéric

  • Author_Institution
    Utah Center for Advanced Imaging Research, Department of Radiology, University of Utah, Salt Lake City, USA
  • fYear
    2008
  • fDate
    19-25 Oct. 2008
  • Firstpage
    4444
  • Lastpage
    4447
  • Abstract
    We consider estimation of channelized Hotelling observer performance measures from a finite sample of reconstructed CT images, resulting from repeated scans. After observing that precise statistical theory for estimates of the SNR and AUC does not appear to be well-known in the medical imaging literature, we demonstrate how known results from the statistics literature may be applied to better understand the distributions of estimates for the SNR and AUC. In addition, we provide exact expressions for the bias and variance of the estimated SNR2 and show how the moments for the estimated AUC may be numerically evaluated, without Monte Carlo simulation.
  • Keywords
    Area measurement; Automatic voltage control; Computed tomography; Distributed computing; Estimation theory; Image reconstruction; Lesions; Nuclear and plasma sciences; Nuclear measurements; Signal to noise ratio;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2008. NSS '08. IEEE
  • Conference_Location
    Dresden, Germany
  • ISSN
    1095-7863
  • Print_ISBN
    978-1-4244-2714-7
  • Electronic_ISBN
    1095-7863
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2008.4774268
  • Filename