DocumentCode :
2615882
Title :
Simulation of mechanical misalignments in a cone-beam micro-CT system
Author :
Vidal-Migallón, I. ; Abella, M. ; Sisniega, A. ; Vaquero, J.J. ; Desco, M.
Author_Institution :
Unidad de Medicina y Cirug?a Experimental, Hospital General Universitario Gregorio Mara??n, Spain
fYear :
2008
fDate :
19-25 Oct. 2008
Firstpage :
5007
Lastpage :
5009
Abstract :
X-ray CT images usually show artefacts due not only to physical effects -e.g., beam hardening-, but also to misalignments that remain after mechanical calibration. These artefacts become particularly noticeable in the case of high spatial resolution systems and in hybrid systems, such as PET-CT, SPECT-CT scanners, which rely on a correct registration of emission and CT data. Hence, slight mechanical misalignments affect the quality of the CT images and any attenuation correction methods or further quantification based on them. We implemented a computer simulator of these artefacts on a cone-beam, flat-panel based micro-CT scanner. Using this simulator, we studied the effect of these different misalignments (pitch and roll tilts, skew and shifts) on reconstructed images.
Keywords :
Attenuation; Calibration; Computational modeling; Computed tomography; Detectors; Image reconstruction; Imaging phantoms; Nuclear and plasma sciences; Projection algorithms; Spatial resolution;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record, 2008. NSS '08. IEEE
Conference_Location :
Dresden, Germany
ISSN :
1095-7863
Print_ISBN :
978-1-4244-2714-7
Electronic_ISBN :
1095-7863
Type :
conf
DOI :
10.1109/NSSMIC.2008.4774364
Filename :
4774364
Link To Document :
بازگشت