Scatter/trues detection efficiency compensation in scatter correction of PET emission data
Author
Watson, Charles ; Byars, Larry ; Michel, Christian ; Rothfuss, Harold
Author_Institution
Siemens Medical Solutions Molecular Imaging, Knoxville, TN 37932 USA
fYear
2008
fDate
19-25 Oct. 2008
Firstpage
5493
Lastpage
5497
Abstract
In positron emission tomography (PET), a detector’s response to scattered radiation may be different from its response to unscattered (true coincidence) photons. This difference should be accounted for during normalization and scatter correction. We show that only a knowledge of the ratio of the scatter to trues efficiencies is necessary, however. We describe a new technique for measuring the scatter/trues detection efficiency ratio, and a procedure for applying this compensation during the scatter correction of PET emission data. We present some initial results, confirmed by simulations, indicating that the scatter/trues efficiencies may vary by up to 20% for a Siemens BrainPET MR insert in the vicinity of block gaps near the center of the transaxial field of view. These variations could result in ±5% variations in scatter corrected images if left uncompensated.
Keywords
Brain modeling; Electromagnetic scattering; Image reconstruction; Nuclear and plasma sciences; Particle scattering; Photonics; Positron emission tomography; Probability distribution; Radiation detectors; Single photon emission computed tomography;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium Conference Record, 2008. NSS '08. IEEE
Conference_Location
Dresden, Germany
ISSN
1095-7863
Print_ISBN
978-1-4244-2714-7
Electronic_ISBN
1095-7863
Type
conf
DOI
10.1109/NSSMIC.2008.4774496
Filename
4774496
Link To Document