DocumentCode
2620491
Title
Test of a 3”×3” LaCl3 :Ce crystal and neutron sensitivity of lanthanum halide scintillators
Author
Martínez, T. ; Cano-Ott, Daniel ; Guerrero, C.
Author_Institution
Department of Energy, CIEMAT, Avda. Complutense, 22 28040 Madrid (Spain)
fYear
2008
fDate
19-25 Oct. 2008
Firstpage
1220
Lastpage
1222
Abstract
The availability of the new scintillator materials LaCl3 and LaBr3 offers a good opportunity for the design of new detector assemblies in different research fields. In particular, the total absorption technique employed in neutron capture measurements could be highly benefit from the use of these two scintillation materials due their excellent energy resolution and fast decay time properties. In this work we have tested a 3”×3” LaCl3 detector prototype coupled to a XP5300B photomultiplier. We have determined experimentally the energy resolution at different γ-ray energies and the intrinsic background related to 138La. In addition we have studied the neutron sensitivity of both materials by means of GEANT4 Monte Carlo simulations.
Keywords
Absorption; Assembly; Energy capture; Energy measurement; Energy resolution; Neutrons; Particle measurements; Solid scintillation detectors; Testing; Time measurement;