• DocumentCode
    2621620
  • Title

    MSGCROC - an ASIC for high count rate readout of position sensitive Microstrip Gas Chambers for thermal neutrons

  • Author

    Brogna, Andrea S. ; Buzzetti, Siro ; Dabrowski, Wladyslaw ; Fiutowski, Tomasz ; Gebauer, Burckhard ; Mindur, Bartosz ; Schmidt, Christian J. ; Schulz, Christian ; Soltveit, Hans K. ; Szczygiel, Robert ; Trunk, Ulrich ; Wiacek, Piotr

  • Author_Institution
    Physikalisches Institut der Universitÿt Heidelberg, D-69120, Germany
  • fYear
    2008
  • fDate
    19-25 Oct. 2008
  • Firstpage
    1482
  • Lastpage
    1488
  • Abstract
    In this paper we report on the development of an ASIC for readout of position sensitive neutron detectors based on low-pressure Micro-Strip Gas Chambers with solid composite 157Gd/CsI converter. Global counting rates of 108 cps for a detector area of 25×25 cm2 covered with 400×400 strips requiring 2 ns time resolution for coincidences of signals from X- and Y-strips set very demanding requirements for the ASIC performance.
  • Keywords
    Application specific integrated circuits; Microstrip; Neutrons; Position sensitive particle detectors; Signal resolution; Silicon; Spatial resolution; Strips; X-ray detection; X-ray detectors;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2008. NSS '08. IEEE
  • Conference_Location
    Dresden, Germany
  • ISSN
    1095-7863
  • Print_ISBN
    978-1-4244-2714-7
  • Electronic_ISBN
    1095-7863
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2008.4774695
  • Filename