Title :
Drift-time calibration of the ATLAS Monitored Drift Tubes with a precision mini-chamber
Author :
Levin, Daniel S. ; Amram, Nir ; ben Moshe, Meni ; Etzion, Erez ; Dai, Tiesheng ; Diehl, Edward B. ; Ferretti, Claudio ; Gregory, Jeffery ; Kiesel, Mike ; Thun, Rudi ; Weaverdyck, Curtis ; Wilson, Alan ; Zhou, Bing
Author_Institution :
Department of Physics, University of Michigan, USA
Abstract :
Muon track reconstruction in the ATLAS Muon Precision (MDT) Chambers relies on many thousands of calibration parameters: T-zeros offsets and on time-to-space (RT) functions that convert drift time to drift radius. A data intensive computing program has been created to obtain compute these constants daily for each of thousands of muon-spectrometer subregions. For early LHC running a simplified calibration employing a single universal RT function (URT) provided by a dedicated gas-monitor MDT chamber can be invoked. Constructed with standard MDT’s, the monitor chamber records cosmic ray drift-time spectra derived from gas flowing in the input and output arteries of the main ATLAS MDT gas supply. These spectra are used to compute the URT bi-hourly, referenced to a standard operating pressure and temperature.
Keywords :
Assembly; Calibration; Large Hadron Collider; Magnetic multilayers; Mesons; Monitoring; Spatial resolution; Spectroscopy; Temperature measurement; Toroidal magnetic fields;
Conference_Titel :
Nuclear Science Symposium Conference Record, 2008. NSS '08. IEEE
Conference_Location :
Dresden, Germany
Print_ISBN :
978-1-4244-2714-7
Electronic_ISBN :
1095-7863
DOI :
10.1109/NSSMIC.2008.4774805