DocumentCode :
2624272
Title :
Use of standard CMOS pixel imagers as ionizing radiation detectors
Author :
Servoli, L. ; Biagetti, D. ; Passeri, D. ; Gattuso, E.Spanti
Author_Institution :
Istituto Nazionale di Fisica Nucleare, Sezione di Perugia, I-06132 Italy
fYear :
2008
fDate :
19-25 Oct. 2008
Firstpage :
2484
Lastpage :
2488
Abstract :
The recent developments in the domain of the standard CMOS imagers for visible light, mainly in the reduction of the pixel size, has led us to investigate the suitability of some of these devices as ionizing radiation detectors. A standard 640x480 imager with 5.6x5.6 micrometer pixel size (Micron product type MT9V011) has been characterized with both photons ( 55Fe and 8 keV X-ray tube) than charged particles sources ( 500 MeV electrons). The main results obtained are: a small pixel multiplicity (4–5), for the detection of either X-ray or charged particles; a good linearity of the response; a S/N ratio ≫ 30 for a Minimum Ionizing Particle and an estimated sensitivity down to about 2 keV.
Keywords :
Electron tubes; Ionizing radiation; Iron; Linearity; Pixel; Radiation detectors; Standards development; X-ray detection; X-ray detectors; X-ray imaging;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record, 2008. NSS '08. IEEE
Conference_Location :
Dresden, Germany
ISSN :
1095-7863
Print_ISBN :
978-1-4244-2714-7
Electronic_ISBN :
1095-7863
Type :
conf
DOI :
10.1109/NSSMIC.2008.4774860
Filename :
4774860
Link To Document :