• DocumentCode
    2624541
  • Title

    Design of a Pill-Sized 12-legged Endoscopic Capsule Robot

  • Author

    Quirini, Marco ; Webster, Robert J., III ; Menciassi, Arianna ; Dario, Paolo

  • Author_Institution
    CREV1 Lab., Scuola Superiore Sant´´Anna, Pisa
  • fYear
    2007
  • fDate
    10-14 April 2007
  • Firstpage
    1856
  • Lastpage
    1862
  • Abstract
    In this paper we present the design of a swallowable (11mm diameter by 25mm long), 12-legged endoscopic capsule for locomotion in the lower gastro intestinal tract (large bowel). A novel slot-follower mechanism driven via lead-screw allows the capsule to be as small as current commercial pill-cameras, while simultaneously generating 2/3 N of force at each leg tip. Kinematic and static analyses of the lead screw and slot-follower mechanisms allow optimization of design parameters so that the capsule satisfies experimental and clinical design requirements for legged locomotion in the GI tract.
  • Keywords
    medical robotics; microrobots; mobile robots; robot kinematics; 11 mm; 25 mm; lead screw; legged locomotion; lower gastro intestinal tract; pill-sized 12-legged endoscopic capsule robot; robot kinematics; slot-follower mechanism; Cancer; Colon; Fasteners; Gastrointestinal tract; Intestines; Leg; Legged locomotion; Magnetic fields; Robotics and automation; Robots;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Robotics and Automation, 2007 IEEE International Conference on
  • Conference_Location
    Roma
  • ISSN
    1050-4729
  • Print_ISBN
    1-4244-0601-3
  • Electronic_ISBN
    1050-4729
  • Type

    conf

  • DOI
    10.1109/ROBOT.2007.363592
  • Filename
    4209356