Title :
24.7 A 60nV/√Hz 15-channel digital active electrode system for portable biopotential signal acquisition
Author :
Jiawei Xu ; Busze, Benjamin ; Hyejung Kim ; Makinwa, Kofi ; Van Hoof, Chris ; Yazicioglu, Refet Firat
Author_Institution :
Holst Centre/imec, Eindhoven, Netherlands
Abstract :
Dry active electrodes (AE), i.e., the combination of dry electrodes with in situ amplification, are increasingly used for biopotential measurements in emerging healthcare and lifestyle applications [1]. Compared to gel-based wet electrodes, dry electrodes enable fast set-up time, greater user comfort, and long-term monitoring. AE amplifiers ensure local amplification providing improved robustness to noise interference and cable motion artifacts. However, current AEs have analog outputs requiring powerful analog buffers to drive biopotential signals over measurement cables. Furthermore, analog outputs must be digitized by the back-end (BE) system [2,3]. Besides, parameter mismatch between AEs limits the overall CMRR. CMFB [1] or CMFF [2] helps but comes at the expense of increased number cables between the BE and AEs. These problems significantly increase the overall system complexity and cost.
Keywords :
amplifiers; analogue-digital conversion; bioelectric potentials; biomedical electrodes; cables (electric); medical signal detection; medical signal processing; patient monitoring; portable instruments; 15-channel digital active electrode system; AE amplifiers; AE analog outputs; AE parameter mismatch; BE system; CMFB; CMFF; CMRR limitation; analog output digitization; back-end system; biopotential measurements; biopotential signal driving; cable motion artifacts; cable number; dry active electrodes; fast set-up time; gel-based wet electrodes; healthcare applications; in situ amplification; lifestyle applications; long-term monitoring; measurement cables; noise interference; portable biopotential signal acquisition; powerful analog buffers; robustness improvement; system complexity; system cost; user comfort; voltage 60 nV; DSL; Electrodes; Electroencephalography; Electronics packaging; Impedance; Monitoring; Noise;
Conference_Titel :
Solid-State Circuits Conference Digest of Technical Papers (ISSCC), 2014 IEEE International
Conference_Location :
San Francisco, CA
Print_ISBN :
978-1-4799-0918-6
DOI :
10.1109/ISSCC.2014.6757498