• DocumentCode
    2625006
  • Title

    Performance of a high speed and high density data acquisition system for multiple gamma-ray detection

  • Author

    Kimura, Atsushi ; Toh, Yosuke ; Koizumi, Mitsuo ; Furutaka, Kazuyoshi ; Kin, Tadahiro ; Oshima, Masumi

  • Author_Institution
    Research group for Innovative nuclear science, Japan Atomic Energy Agency, Tokai-mura, Ibaraki-ken, JAPAN
  • fYear
    2008
  • fDate
    19-25 Oct. 2008
  • Firstpage
    2107
  • Lastpage
    2111
  • Abstract
    To accelerate the data acquisition system (DAQ), which was reported in NSS 2004[1], the authors improved the Main ADC Modules in the DAQ. The new Main ADC Module has 8 input channels. Each channel consists of analog circuits for pulse shaping, a high-speed ADC (14bit 125MSPS), and a highpower Digital Signal Processor (DSP: 8000MIPS). Shaped input signals are digitized with the ADC and analyzed by the DSP to calculate the energy data. Total energy resolution of this system was 2.5 keV in FWHM for 1.33-MeV gamma rays. Time resolution of this system was 120 ns in FWHM between two Ge detectors. These results show good performances as a conventional DAQ. The dead time of this system is only 4.3 µs per event and the maximum event rate was more than 350k events/s. Thus, enough energy and time resolutions are achieved with short dead time, high channel density and high event rate.
  • Keywords
    Acceleration; Analog circuits; Data acquisition; Digital signal processing; Digital signal processors; Energy resolution; Gamma ray detection; Gamma ray detectors; Pulse shaping methods; Signal analysis;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2008. NSS '08. IEEE
  • Conference_Location
    Dresden, Germany
  • ISSN
    1095-7863
  • Print_ISBN
    978-1-4244-2714-7
  • Electronic_ISBN
    1095-7863
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2008.4774912
  • Filename
    4774912