• DocumentCode
    2628246
  • Title

    Geant4 validation on mammography applications

  • Author

    Vidal, Feijó Pery ; Gabriela, Hoff

  • Author_Institution
    Grupo de Simulação Computacional em FÃ\xadsica Médica, Universidade Católica do Rio Grande do Sul, Av Ipiranga 6681, Partenon, Porto Alegre, CEP 90619-900, RS, Brazil
  • fYear
    2008
  • fDate
    19-25 Oct. 2008
  • Firstpage
    3497
  • Lastpage
    3498
  • Abstract
    The objective of this work is to validate the Geant4 for mammographic applications. The computational geometry simulates a LORAD MIII mammographic equipment, operating into the range of 24 kVp up to 34 kVp with target-add filtration material and thickness Mo-30μm Mo and Mo-25μm Rh. Were compared values of normalized glandular dose (DgN) and backscattering factor (BSF). Were simulated non-antropomorphic models composed by a homogeneous distribution of BR12 and PMMA of dimensions (10 × 12 × 4) cm3 and (10 × 12 × 8) cm3. We present the comparison of BSF of our simulation to Kramer, EUR report and Hoff published data. The results presented on this work can be considered a good agreement The validation using absolute dosimetric quantities like exposure and KERMA can be done, but it is important to define the number de photons emitted by mAs. We suggest the use of a semi-empirical correction factor based on tube enhancement test to realize this kind of validation.
  • Keywords
    Backscatter; Computational geometry; Computational modeling; Equations; Filtration; Mammography; Milling machines; Nuclear and plasma sciences; Solid modeling; Testing;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2008. NSS '08. IEEE
  • Conference_Location
    Dresden, Germany
  • ISSN
    1095-7863
  • Print_ISBN
    978-1-4244-2714-7
  • Electronic_ISBN
    1095-7863
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2008.4775089
  • Filename
    4775089