Title :
Towards fully automated precise measurement of camera transfer functions
Author :
Friedrich, David ; Brauers, Johannes ; Bell, André A. ; Aach, Til
Author_Institution :
Inst. of Imaging & Comput. Vision, RWTH Aachen Univ., Aachen, Germany
Abstract :
The mapping of incident irradiance to sensor output value, the so-called camera transfer function (CTF), of CCD or CMOS imaging devices is in general a non-linear function. Applications that utilize the camera as a measurement device for the incident irradiance, e.g DNA image cytometry for early cancer detection or HDR imaging, require a linear CTF. Thus, this non-linearity has to be corrected. This, in turn, requires the non-linear CTF to be estimated or measured. In this paper we present a radiometric measurement setup. We point out a particular problem, i.e., the calibration of the distance offset between sensor and light source. We then show a solution which integrates this calibration into the measurement process, enabling a fully automated measurement of the CTF.
Keywords :
CMOS image sensors; Calibration; Cameras; Cancer detection; Charge coupled devices; Charge-coupled image sensors; DNA; Light sources; Radiometry; Transfer functions; CCD; CMOS; Camera transfer function; Non-linearity; Opto electrical conversion function; Radiometric calibration;
Conference_Titel :
Image Analysis & Interpretation (SSIAI), 2010 IEEE Southwest Symposium on
Conference_Location :
Austin, TX, USA
Print_ISBN :
978-1-4244-7801-9
DOI :
10.1109/SSIAI.2010.5483897