• DocumentCode
    2639000
  • Title

    Experimental investigation on calibrating high precision accelerometer on two-axis table

  • Author

    Xiaoxiong, Jiang ; Yu, Liu ; Baoku, Su ; Ming, Liu

  • Author_Institution
    Space Control & Inertial Technol. Res. Center, Harbin Inst. of Technol., Harbin
  • fYear
    2008
  • fDate
    10-12 Dec. 2008
  • Firstpage
    1
  • Lastpage
    5
  • Abstract
    Model identification of accelerometer is one of the most pervasive problems in calibration test of accelerometer. A method of calibrating dual orthogonal accelerometers on two-axis table is presented to eliminate the effect of the angle-setting error on calibration, thus enhancing identification precision of the accelerometer. The correlation of the model coefficients of dual orthogonal accelerometers is analyzed and the D-optimal experiment design is accomplished. Furthermore analytical solutions are deduced. The simulation experiment and measured data suggests that the standard deviation of identification of dual orthogonal accelerometers is one order of magnitude less than that of single accelerometer. Therefore, the precision of identifying the error model of accelerometer is improved.
  • Keywords
    accelerometers; calibration; design of experiments; D-optimal experiment design; angle-setting error; dual orthogonal accelerometers; high precision accelerometer calibration; model identification; two-axis table; Accelerometers; Calibration; Measurement standards; Testing;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Systems and Control in Aerospace and Astronautics, 2008. ISSCAA 2008. 2nd International Symposium on
  • Conference_Location
    Shenzhen
  • Print_ISBN
    978-1-4244-3908-9
  • Electronic_ISBN
    978-1-4244-2386-6
  • Type

    conf

  • DOI
    10.1109/ISSCAA.2008.4776348
  • Filename
    4776348