DocumentCode
2639000
Title
Experimental investigation on calibrating high precision accelerometer on two-axis table
Author
Xiaoxiong, Jiang ; Yu, Liu ; Baoku, Su ; Ming, Liu
Author_Institution
Space Control & Inertial Technol. Res. Center, Harbin Inst. of Technol., Harbin
fYear
2008
fDate
10-12 Dec. 2008
Firstpage
1
Lastpage
5
Abstract
Model identification of accelerometer is one of the most pervasive problems in calibration test of accelerometer. A method of calibrating dual orthogonal accelerometers on two-axis table is presented to eliminate the effect of the angle-setting error on calibration, thus enhancing identification precision of the accelerometer. The correlation of the model coefficients of dual orthogonal accelerometers is analyzed and the D-optimal experiment design is accomplished. Furthermore analytical solutions are deduced. The simulation experiment and measured data suggests that the standard deviation of identification of dual orthogonal accelerometers is one order of magnitude less than that of single accelerometer. Therefore, the precision of identifying the error model of accelerometer is improved.
Keywords
accelerometers; calibration; design of experiments; D-optimal experiment design; angle-setting error; dual orthogonal accelerometers; high precision accelerometer calibration; model identification; two-axis table; Accelerometers; Calibration; Measurement standards; Testing;
fLanguage
English
Publisher
ieee
Conference_Titel
Systems and Control in Aerospace and Astronautics, 2008. ISSCAA 2008. 2nd International Symposium on
Conference_Location
Shenzhen
Print_ISBN
978-1-4244-3908-9
Electronic_ISBN
978-1-4244-2386-6
Type
conf
DOI
10.1109/ISSCAA.2008.4776348
Filename
4776348
Link To Document