• DocumentCode
    2653134
  • Title

    Dynamic olfactometer used as a professional “nose”

  • Author

    Liu, Weiling ; Bao, Jingling ; Zou, Kehua ; Li, Changjian ; Xu, Yan ; Wang, Liyuan

  • Author_Institution
    State Environ. Protection Key Lab. of Odor, Tianjin Acad. of Environ. Sci., Tianjin, China
  • fYear
    2010
  • fDate
    14-18 Dec. 2010
  • Firstpage
    1327
  • Lastpage
    1330
  • Abstract
    The aim of this work is to establish a Dynamic olfactometer used as a professional nose for odor concentration measurement. The study was composed of three proposals: establishing a dynamic dilution olfactometer, calibration the olfactometer, and screening panelist. In this paper, we provide detailed instructions for the pneumatic system of the olfactometer. Criteria were used in an odor test following EN13725 and GB T14675-93 standard. Oder concentration was determined using triangle forced-choice method. The computerized dynamic olfactometer can realize with 13 dilution levels and dilution ratios 10 to 105. The central part is mass flow meter, integrated LabVIEW platform with Microsoft® Access database. Result revealed that the olfactometer can achieve an averaged instability of 2.5% and an averaged accuracy of 3% over the dilution range of 10-105.
  • Keywords
    chemical variables measurement; electronic noses; calibration; dynamic dilution olfactometer; mass flow meter; odor concentration measurement; odor test; pneumatic system; Accuracy; Europe; Manifolds; Nose; Pollution measurement; Software; Valves; Dynamic Dilution; odor concentration; olfactormeter; triangle forced-choice;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Robotics and Biomimetics (ROBIO), 2010 IEEE International Conference on
  • Conference_Location
    Tianjin
  • Print_ISBN
    978-1-4244-9319-7
  • Type

    conf

  • DOI
    10.1109/ROBIO.2010.5723521
  • Filename
    5723521