DocumentCode
2657253
Title
Cavity pulling in Galileo passive hydrogen maser
Author
Busca, G. ; Wang, Q. ; Belloni, M. ; Mattioni, L.
fYear
2003
fDate
4-8 May 2003
Firstpage
86
Lastpage
89
Abstract
The paper derives the theoretical expression of the cavity pulling factor in Galileo space passive hydrogen Maser, based on the ´single frequency modulation´. The sensitivity of the cavity pulling vs. the phase selection is deduced in order to account for the experimental errors in the phase adjustments. The theoretical calculation shows the new phase setting exists to achieve a zero cavity pulling, and the preliminary experimental results comparing with the two phase selection are presented.
Keywords
frequency modulation; hydrogen; masers; Galileo space passive hydrogen Maser; H; cavity pulling factor; frequency modulation; phase adjustments; zero cavity pulling; Aerospace electronics; Envelope detectors; Frequency modulation; Hydrogen; Masers; Phase detection; Phase modulation; Prognostics and health management; Resonance; Resonant frequency;
fLanguage
English
Publisher
ieee
Conference_Titel
Frequency Control Symposium and PDA Exhibition Jointly with the 17th European Frequency and Time Forum, 2003. Proceedings of the 2003 IEEE International
ISSN
1075-6787
Print_ISBN
0-7803-7688-9
Type
conf
DOI
10.1109/FREQ.2003.1274992
Filename
1274992