Title :
IEN-CsF1 accuracy evaluation and two way frequency comparison
Author :
Levi, F. ; Lorini, L. ; Calonico, D. ; Bertacco, E.K. ; Godone, A.
Author_Institution :
Ist. Elettrotecnico Nazionale "G. Ferraris", Torino, Italy
Abstract :
In this paper we report the accuracy evaluation of the Italian primary frequency standard IEN-CsF1. We discuss the frequency shifts our frequency standard is corrected for and the procedure used for accuracy evaluation. In the last section we report a frequency comparison against two remote laboratory, the National Institute of Standard and Technology and the Physikalisch-Technische Bundesanstalt using the two way comparison technique.
Keywords :
atomic clocks; caesium; frequency stability; frequency standards; laboratories; Cs; Italian primary frequency standard IEN-CsF1; National Institute of Standard and Technology; Physikalisch-Technische Bundesanstalt; accuracy evaluation; frequency comparison; frequency shift; frequency standard; remote laboratory; Atomic beams; Atomic measurements; Clocks; Electron tubes; Frequency measurement; Laser beams; Laser modes; Masers; Measurement standards; Time measurement;
Conference_Titel :
Frequency Control Symposium and PDA Exhibition Jointly with the 17th European Frequency and Time Forum, 2003. Proceedings of the 2003 IEEE International
Print_ISBN :
0-7803-7688-9
DOI :
10.1109/FREQ.2003.1275087