Title :
Improvement in accuracy of a single 40Ca+ optical clock towards the 10−15 level using a magnetic shield
Author :
Matsubara, Kensuke ; Li, Ying ; Nagano, Shigeo ; Ito, Hiroyuki ; Kajita, Masatoshi ; Kojima, Reiko ; Hanado, Yuko ; Hayasaka, Kazuhiro ; Hosokawa, Mizuhiko
Author_Institution :
Space-Time Stand. Group, Nat. Inst. of Inf. & Commun., Tokyo, Japan
Abstract :
We report a frequency measurement of a single 40Ca+ optical clock towards the 10-15 uncertainty level. In contrast with our previous measurement, in which the accuracy was limited by line broadening due to the magnetic field fluctuation, the use of a magnetic shield significantly reduced the spectrum linewidth of the clock transition. By a frequency comparison to an 87Sr lattice clock, an Allan deviation of 7 × 10-16 was observed at an averaging time of 103 seconds. We preliminarily measured the clock transition frequency to be 411 042 129 776 397 (3) Hz with respect to the SI definition of the second. Further measurements are executed aiming at an accuracy of low parts in the 10-15 level.
Keywords :
atomic clocks; frequency measurement; magnetic fields; magnetic shielding; Allan deviation; clock transition frequency measurement; line broadening; magnetic field fluctuation; magnetic shield; optical clock; spectrum linewidth; Clocks; Frequency measurement; Laser stability; Laser transitions; Measurement by laser beam; Optical variables measurement; Uncertainty;
Conference_Titel :
Frequency Control and the European Frequency and Time Forum (FCS), 2011 Joint Conference of the IEEE International
Conference_Location :
San Fransisco, CA
Print_ISBN :
978-1-61284-111-3
DOI :
10.1109/FCS.2011.5977880