Title :
A High-Power Dual Six-Port Automatic Network Analyzer for Determining Biological Effects of RF and Microwave Radiation
Abstract :
The design, calibration and performance of a high-power (1 to 1000W) automatic network analyzer based on the six-port concept are described for the 10 to 100 MHz range. Imprecision in measuring reflection coefficient Gamma is 0.0001 in magnitude and 0.005/ |Gamma| degrees in phase. Corresponding estimated systematic errors are 0.001 and 0.1/ |Gamma| degrees. Imprecision in measuring power is 0. 01% of range (20W, 200W, or 1000W) with an estimated systematic error of 1. 25% of reading.
Keywords :
Calibration; Electromagnetic radiation; NIST; Phase measurement; Power measurement; Radio frequency; Reflection; Switches; Synthesizers; Testing;
Conference_Titel :
Microwave Symposium Digest, 1981 IEEE MTT-S International
Conference_Location :
Los Angeles, CA, USA
DOI :
10.1109/MWSYM.1981.1129854