Title :
Performance measurements on active cold loads for radiometer calibration
Author :
Skou, Niels ; Søbjærg, Sten S. ; Balling, Jan
Author_Institution :
B 348 Tech. Univ. of Denmark, Lyngby
Abstract :
Two semi-conductor Active Cold Loads (ACLs) to be used as cold references in spaceborne microwave radiometers have been developed. An X-band frequency has been chosen, and the target noise temperature value is in the 50 to 100 K range. The ACLs are to be characterized in the operating temperature range 0-50degC, and long term stability must be assessed. To this end a Test Bed has been developed. This Test Bed is actually a stable radiometer, and its design and development is discussed. The test setup is described, and preliminary test campaign results indicate output temperatures of 73 K and 51 K for the two ACLs. The temperature sensitivity is slightly below 0.4 K/degC for the units, and excellent stability seems to prevail.
Keywords :
calibration; radiometers; semiconductor device testing; performance measurements; radiometer calibration; semiconductor active cold loads; spaceborne microwave radiometers; temperature 0 degC to 50 degC; temperature 50 K to 100 K; temperature sensitivity; Calibration; Extraterrestrial measurements; Frequency; Microwave radiometry; Space cooling; Stability; System testing; Temperature distribution; Temperature measurement; Temperature sensors; active cold load; calibration; microwave; radiometer;
Conference_Titel :
Geoscience and Remote Sensing Symposium, 2007. IGARSS 2007. IEEE International
Conference_Location :
Barcelona
Print_ISBN :
978-1-4244-1211-2
Electronic_ISBN :
978-1-4244-1212-9
DOI :
10.1109/IGARSS.2007.4423333