Title :
Gamma Radiation Sensing Using ZnO and SnO2 Thick Film Interdigitated Capacitors
Author :
Arshak, K. ; Korostynska, O. ; Jafer, E. ; Arshak, A. ; Morris, D. ; Gill, E.
Author_Institution :
Univ. of Limerick, Limerick
Abstract :
A novel cost-effective real-time gamma radiation monitoring system based on metal oxide thick films was designed and tested. The changes in capacitances of ZnO and SnO2 thick film capacitors with interdigitated electrodes were monitored in real time under the influence of gamma radiation using miniaturized, low power, bi-directional wireless communication system. The capacitive interface circuitry was based on a Delta-sigma (SigmaDelta) modulator using switched capacitor (SC) circuit architecture with integrated on-chip temperature sensor. At the base station side, an frequency shift keying (FSK) receiver/transmitter is connected to another MCU unit, which send the received data or received instructions from a PC through a graphical user interface GUI. Industrial, scientific and medical (ISM) band RF (433 MHz) was used to achieve half duplex communication between the two sides. All the modules of the mixed signal system are integrated in a printed circuit board (PCB) of size 22.46times20.168 mm. The overall system supply voltage is 2.7 V maximum. 137Cs source with an activity of 370 kBq was used. An increase in the values of capacitance with radiation was recorded for both films to a certain dose level, determined by the material of sensitive layer.
Keywords :
dosimetry; gamma-ray detection; interdigital transducers; radiotelemetry; thick film capacitors; thick film sensors; tin compounds; zinc compounds; ISM band; PCB; SnO2; ZnO; bi-directional wireless communication system; capacitive interface circuitry; delta-sigma modulator; dose level; frequency shift keying receiver; frequency shift keying transmitter; gamma radiation sensing; half duplex communication; integrated on-chip temperature sensor; interdigitated electrodes; metal oxide thick films; mixed signal system; printed circuit board; radioactivity 370 kBq; real-time gamma radiation monitoring system; switched capacitor circuit architecture; thick film interdigitated capacitors; Capacitance; Capacitors; Circuits; Frequency shift keying; Gamma rays; Graphical user interfaces; Radiation monitoring; Real time systems; Thick films; Zinc oxide;
Conference_Titel :
Electronics Technology, 2006. ISSE '06. 29th International Spring Seminar on
Conference_Location :
St. Marienthal
Print_ISBN :
1-4244-0551-3
Electronic_ISBN :
1-4244-0551-3
DOI :
10.1109/ISSE.2006.365154