DocumentCode
2677191
Title
2D imaging with a single-sided MPI device
Author
Grafe, K. ; von Gladiss, A. ; Bringout, G. ; Ahlborg, M. ; Buzug, T.M.
Author_Institution
Inst. of Med. Eng., Univ. zu Lubeck, Lubeck, Germany
fYear
2015
fDate
26-28 March 2015
Firstpage
1
Lastpage
1
Abstract
This paper presents 2D images that have been measured successfully with a single-sided magnetic particle imaging (MPI) scanning device. This provides proof that multidimensional single-sided MPI is possible. At the moment, the highest achieved penetration depth is about 10 mm. The penetration depth is limited by the inhomogeneous magnetic fields, which are caused by the asymmetric coil assembly. Besides, a measurement width of 30 mm is possible in y-direction at the cost of a reduced accuracy. The next steps aim to improve the signal to noise ratio, to calculate the resolution and the sensitivity of the single-sided scanner.
Keywords
biomedical MRI; 2D imaging; asymmetric coil assembly; inhomogeneous magnetic fields; multidimensional single-sided MPI; penetration depth; scanner resolution; sensitivity; signal-to-noise ratio; single-sided MPI device; single-sided magnetic particle imaging; Coils; Image reconstruction; Imaging; Magnetic field measurement; Magnetic fields; Magnetic particles; Trajectory;
fLanguage
English
Publisher
ieee
Conference_Titel
Magnetic Particle Imaging (IWMPI), 2015 5th International Workshop on
Conference_Location
Istanbul
Print_ISBN
978-1-4799-7269-2
Type
conf
DOI
10.1109/IWMPI.2015.7107024
Filename
7107024