• DocumentCode
    2679337
  • Title

    A precision frequency standard comparison method and instrument

  • Author

    Wei Zhou ; Jianguo Yu ; Li, Jie ; Xuan, Zongqiang ; Hui Zhou

  • Author_Institution
    Dept. of Meas. & Instrum., Xidian Univ., Xi´´an, China
  • fYear
    2000
  • fDate
    2000
  • Firstpage
    557
  • Lastpage
    560
  • Abstract
    The linear phase comparison method is used widely in the long-term performance measurement of frequency standards. Combined with a high-speed voltage measuring method, high-resolution short-term stability measurement can also be accomplished based on the phase comparison device. In this case, sampling speed, voltage measuring accuracy and the measuring range will influence the comparison precision of the frequency standard. For higher resolution and convenient use, a “voltage self-correction” method based on linear phase comparison can be used. In the frequency standard comparison, 1×10 -13/τ measurement precision can be obtained and it has a good identical measuring result with other high-precision frequency standard comparators
  • Keywords
    frequency stability; frequency standards; phase comparators; phase measurement; high-resolution stability measurement; high-speed voltage measuring method; linear phase comparison method; measuring range; precision frequency standard comparison method; sampling speed; short-term stability measurement; voltage measuring accuracy; voltage self-correction method; Fluctuations; Frequency measurement; Instruments; Linearity; Measurement standards; Noise measurement; Oscillators; Phase measurement; Stability; Voltage measurement;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Frequency Control Symposium and Exhibition, 2000. Proceedings of the 2000 IEEE/EIA International
  • Conference_Location
    Kansas City, MO
  • ISSN
    1075-6787
  • Print_ISBN
    0-7803-5838-4
  • Type

    conf

  • DOI
    10.1109/FREQ.2000.887417
  • Filename
    887417