DocumentCode :
267960
Title :
Annealing effects on flexible multi-layered parylene-based sensors
Author :
Kim, B.J. ; Washabaugh, E. Peter ; Meng, Ellis
Author_Institution :
Dept. of Biomed. Eng., Univ. of Southern California, Los Angeles, CA, USA
fYear :
2014
fDate :
26-30 Jan. 2014
Firstpage :
825
Lastpage :
828
Abstract :
To mitigate long term, soaking-induced delamination failure of multi-layered Parylene C devices, a post-process annealing step can be employed to increase adhesion between the Parylene layers. However, it has been shown that annealing of Parylene thin films can alter the bulk properties of the polymer, and thus impact final device performance. To elucidate these effects, the mechanical and electrochemical properties, and sensing performance of untreated and annealed Parylene C-platinum electrochemical impedance-based force sensors were compared. Annealing reduced the height (~3%) and increased the stiffness of the Parylene C sensing channel structure (~1.6x), affecting the sensor´s mechanical response. Furthermore, the electrode surface was smoothed as built-in residual stresses were removed during annealing, altering the sensor´s electrochemical properties. Together, these phenomena resulted in a 24% reduction in sensor sensitivity. These results indicate that heat-based effects cannot be ignored for Parylene-metal device systems, including neural microelectrode implants, and that mechanical and electrochemical properties and performance must be determined after heat treatment, such as annealing and sterilization.
Keywords :
adhesion; annealing; delamination; electrochemical sensors; failure analysis; flexible electronics; force sensors; internal stresses; microelectrodes; platinum; polymers; thin film sensors; Pt; adhesion; annealed parylene C-platinum electrochemical impedance-based force sensors; annealing effects; electrochemical property; electrode surface; flexible multilayered parylene-based sensors; heat treatment; heat-based effects; long term soaking-induced delamination failure mitigation; mechanical property; multilayered parylene C devices; neural microelectrode implants; parylene C sensing channel structure; parylene thin films; polymer; post-process annealing step; residual stresses; sensor mechanical response; sensor sensitivity reduction; sterilization; Annealing; Electrodes; Mechanical sensors; Microchannel; Performance evaluation; Surface impedance;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Micro Electro Mechanical Systems (MEMS), 2014 IEEE 27th International Conference on
Conference_Location :
San Francisco, CA
Type :
conf
DOI :
10.1109/MEMSYS.2014.6765768
Filename :
6765768