• DocumentCode
    2683549
  • Title

    Performance comparison between TOA- and RSSI-based localization methods for wireless capsule endoscopy systems

  • Author

    Ito, Takahiro ; Anzai, Daisuke ; Jianqing Wang

  • Author_Institution
    Grad. Sch. of Eng., Nagoya Inst. of Technol., Nagoya, Japan
  • fYear
    2015
  • fDate
    24-26 March 2015
  • Firstpage
    139
  • Lastpage
    143
  • Abstract
    A wireless capsule endoscope (WCE) is a promising medical device for diagnosis of digestive organs, which takes pictures inside digestive organs, such as large and small intestines and transmits them through implant wireless communications. In a realistic WCE scenario, it is important to precisely estimate the WCE location. Generally, the superiority of TOA-based localization over RSSI-based localization has been pointed out in many papers in wireless communication systems. However, because the propagation speed inside of a human body varies due to the influence of its biological tissues (namely, the relative permittivity of human tissues), the realization of precise TOA-based localization needs to accurately estimate the propagation velocity of implant communication signals. So, in this paper, we propose a TOA-based method for joint location/relative permittivity estimation, and then compare the performances between TOA- and RSSI-based WCE location estimation methods. Our computer simulation results reveal that, the TOA-based localization with accurate relative permittivity information outperforms the RSSI-based localization in terms of the localization accuracy, whereas the RSSI-based localization is advantageous if accurate relative permittivity is not obtained. Additionally, the proposed TOA-based localization can achieve not only better localization accuracy than that of the RSSI-based localization but also accurate averaged relative permittivity estimation.
  • Keywords
    RSSI; body area networks; endoscopes; permittivity; time-of-arrival estimation; RSSI-based localization method; TOA-based localization method; location permittivity estimation; relative permittivity estimation; relative permittivity information; wireless capsule endoscopy system; Accuracy; Computational modeling; Estimation; Implants; Permittivity; Receivers; Wireless communication;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Medical Information and Communication Technology (ISMICT), 2015 9th International Symposium on
  • Conference_Location
    Kamakura
  • Type

    conf

  • DOI
    10.1109/ISMICT.2015.7107515
  • Filename
    7107515