    Plane wave medical ultrasound imaging using adaptive beamforming

  • Author

    Holfort, Iben Kraglund ; Gran, Fredrik ; Jensen, Jørgen Arendt

  • Author_Institution
    Dept. of Electr. Eng., Tech. Univ. of Denmark, Lyngby
  • fYear
    2008
  • fDate
    21-23 July 2008
  • Firstpage
    288
  • Lastpage
    292
  • Abstract
    In this paper, the adaptive, minimum variance (MV) beam-former is applied to medical ultrasound imaging. The significant resolution and contrast gain provided by the adaptive, minimum variance (MV) beamformer, introduces the possibility of plane wave (PW) ultrasound imaging. Data is obtained using Field II and a 7 MHz, 128-elements, linear array transducer with lambda/2-spacing. MV is compared to the conventional delay-and-sum (DS) beamformer with Boxcar and Hanning weights. Furthermore, the PW images are compared to the a conventional ultrasound image, obtained from a linear scan sequence. The four approaches, {Linear Scan, DS Boxcar, DS Hanning, MV}, have full width at half maximum of {0.82, 0.71, 1.28, 0.12} mm and peak side-lobe levels of {-40.1,-16.8,-34.4,-57.0} dB.
  • Keywords
    array signal processing; biomedical imaging; Boxcar weights; Field II; Hanning weights; adaptive beamformer; adaptive beamforming; delay-and-sum beamformer; frequency 7 MHz; linear array transducer; linear scan sequence; medical ultrasound imaging; minimum variance beamformer; plane wave ultrasound imaging; Adaptive arrays; Array signal processing; Biomedical imaging; Biomedical transducers; Focusing; Image resolution; Image sensors; Sensor arrays; Ultrasonic imaging; Ultrasonic transducers;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Sensor Array and Multichannel Signal Processing Workshop, 2008. SAM 2008. 5th IEEE
  • Conference_Location
    Darmstadt
  • Print_ISBN
    978-1-4244-2240-1
  • Electronic_ISBN
    978-1-4244-2241-8
  • Type

    conf

  • DOI
    10.1109/SAM.2008.4606874
  • Filename
    4606874