DocumentCode :
2685786
Title :
Tissue harmonic imaging based on amplitude modulation by controlling number of driving elements
Author :
Ishihara, C. ; Tanaka, Hiroya ; Hashiba, Kunia
Author_Institution :
Central Res. Lab., Hitachi, Ltd., Tokyo, Japan
fYear :
2012
fDate :
7-10 Oct. 2012
Firstpage :
378
Lastpage :
381
Abstract :
A new tissue harmonic imaging (THI) technique, which does not require linearity of a transmission system, based on amplitude modulation by controlling the number of driving elements (AM-CNDE) is proposed. Whether AM-CNDE can achieve THI was investigated by simulation and experiment. It was found out that AM-CNDE can use both harmonics and a fundamental band component although conventional TID (pulse inversion) uses only harmonics to generate images. Due to the fundamental band component, the obtained signal has a broader band than that in the case of pulse inversion. It was confirmed that AM-CNDE achieves TID with almost the same resolution as that by pulse inversion. AM-CNDE is thus considered to be a promising candidate for THI.
Keywords :
amplitude modulation; biological tissues; biomedical ultrasonics; harmonic generation; inverse problems; AM-CNDE; amplitude modulation by controlling the number of driving elements; conventional TID; fundamental band component; pulse inversion; tissue harmonic imaging technique; transmission system; Acoustics; Attenuation; Frequency modulation; Gratings; Image resolution; Time-frequency analysis;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Ultrasonics Symposium (IUS), 2012 IEEE International
Conference_Location :
Dresden
ISSN :
1948-5719
Print_ISBN :
978-1-4673-4561-3
Type :
conf
DOI :
10.1109/ULTSYM.2012.0093
Filename :
6561912
Link To Document :
بازگشت