DocumentCode
2687346
Title
A PC-based wattmeter for high accuracy power measurements
Author
Cataliotti, Antonio ; Cosentino, Valentina ; Di Cara, Dario ; Lipari, Alessandro ; Nuccio, Salvatore ; Spataro, Ciro
Author_Institution
Dept. of Electr., Electron. & Telecommun. Eng. (DIEET), Univ. di Palermo, Palermo, Italy
fYear
2010
fDate
3-6 May 2010
Firstpage
1453
Lastpage
1458
Abstract
This paper proposes the use of a commercial data acquisition board (DAQ) connected to a common personal computer for high accuracy power measurements. In the paper the set up of the PC-based wattmeter is described and an experimental evaluation of its metrological performances is made by means of the comparison with a simultaneous coherent digital sampling wattmeter realized with two high-precision DMMs Hewlett Packard Model 3458 A. In both cases the power source is a Fluke 6100A/E electrical power standard.
Keywords
computerised instrumentation; data acquisition; power measurement; wattmeters; Fluke 6100A/E electrical power standard; Hewlett Packard Model 3458; PC based wattmeter; coherent digital sampling wattmeter; data acquisition board; power measurement; Data acquisition; Instruments; Microcomputers; Performance evaluation; Power measurement; Sampling methods; Signal processing; Transducers; Voltage; Wattmeters; data acquisition board; digital sampling wattmeters; electrical power measurement; pc-based instrument;
fLanguage
English
Publisher
ieee
Conference_Titel
Instrumentation and Measurement Technology Conference (I2MTC), 2010 IEEE
Conference_Location
Austin, TX
ISSN
1091-5281
Print_ISBN
978-1-4244-2832-8
Electronic_ISBN
1091-5281
Type
conf
DOI
10.1109/IMTC.2010.5488096
Filename
5488096
Link To Document