Title :
Ultrasound tissue motion assessment using full correlation analysis
Author :
Taki, Hirofumi ; Sakamoto, Takanori ; Yamakawa, Makoto ; Shiina, Tsuyoshi ; Sato, Takao
Author_Institution :
Kyoto Univ., Kyoto, Japan
Abstract :
Conventional ultrasound cross-correlation technique for tissue motion assessment searches the best match between successive B-mode frames. This strategy neglects the decrease of cross-correlation caused by the time difference between two measurement points utilized for the calculation of cross-correlation. The assumption causes the smaller estimated velocity than the true velocity. In this study, we employ full correlation analysis to compensate for the change of the location where the correlation coefficient is maximum. Simulation study shows that the difference between the true tissue velocity and the expectation of the tissue velocity estimated using the proposed method is 11.3% of that estimated using the conventional cross-correlation technique. These findings indicate the potential of the proposed method to improve the accuracy in tissue motion assessment, including blood flow velocity estimation.
Keywords :
biological tissues; biomedical ultrasonics; correlation methods; blood flow velocity estimation; correlation coefficient; full correlation analysis; simulation study; time difference; tissue velocity estimation expectation; true tissue velocity; ultrasound cross-correlation technique; ultrasound tissue motion assessment; Correlation; Correlation coefficient; Estimation; Time measurement; Ultrasonic imaging; Ultrasonic variables measurement; Velocity measurement; Tissue velocity estimation; cross-correlation; full correlation analysis; ultrasound imaging;
Conference_Titel :
Ultrasonics Symposium (IUS), 2012 IEEE International
Conference_Location :
Dresden
Print_ISBN :
978-1-4673-4561-3
DOI :
10.1109/ULTSYM.2012.0642