DocumentCode :
2690704
Title :
Flexible dry electrode for recording surface electromyogram
Author :
Chan, Adrian D C ; Lemaire, Edward D.
Author_Institution :
Dept. of Syst. & Comput. Eng., Carleton Univ., Ottawa, ON, Canada
fYear :
2010
fDate :
3-6 May 2010
Firstpage :
1234
Lastpage :
1237
Abstract :
A new type of flexible dry electrode is examined for its suitable for surface electromyography (SEMG). SEMG signals were collected from the both biceps of a subject, using the dry electrodes and standard AgAgCl electrodes, during three tasks: 1) rest, 2) an isometric contraction, and 3) a dynamic contraction. Signal quality indices (signal-to-motion artifact ratio, maximum-to-minimum drop in power ratio, signal to noise ratio, and power spectrum deformation) were computed to assess the SEMG. Results show that the dry electrodes can acquire SEMG signals that are nearly indistinguishable from the SEMG signals acquired using the AgAgCl electrodes. The dry electrodes did appear to exhibit a slightly high susceptibility to motion artifact; however, the motion artifact remained below 10 Hz, which can be filtered out for most SEMG applications.
Keywords :
electrodes; electromyography; flexible electronics; flexible dry electrode; surface electromyogram; surface electromyography; Biomedical electrodes; Elbow; Electromyography; Fatigue; Mathematical model; Monitoring; Performance analysis; Root mean square; Signal to noise ratio; Systems engineering and theory; electrodes; electromyography; motion artifact; myoelectric signals;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Instrumentation and Measurement Technology Conference (I2MTC), 2010 IEEE
Conference_Location :
Austin, TX
ISSN :
1091-5281
Print_ISBN :
978-1-4244-2832-8
Electronic_ISBN :
1091-5281
Type :
conf
DOI :
10.1109/IMTC.2010.5488293
Filename :
5488293
Link To Document :
بازگشت