DocumentCode :
2694688
Title :
Detection of HIFU-induced coagulation by motion-compensated decorrelation of RF echo signals
Author :
Umernura, Shin-ichiro ; Matsuzawa, Ryo ; Shishitani, Takashi ; Yoshizawa, Shingo
Author_Institution :
Tohoku Univ., Sendai, Japan
fYear :
2012
fDate :
7-10 Oct. 2012
Firstpage :
2611
Lastpage :
2614
Abstract :
Correlation between the ultrasonic echoes has been used to detect the changes of the tissue during high-intensity focused ultrasound (HIFU) exposure, but it tends be lost when tissue coagulation has progressed. In this study, the loss of correlation itself is utilized to detect the tissue coagulation while compensating the local tissue motion. In the experiment with freshly excised and degassed porcine Iiver, localized decrease in correlation was observed at and around the HIFU focus. The area of decorrelation between each adjacent two frames, separated about 1 s, agreed weil with the actual lesion.
Keywords :
biological tissues; biomedical ultrasonics; coagulation; liver; medical image processing; motion compensation; ultrasonic imaging; HIFU -induced coagulation detection; RF echo signals; degassed porcine Iiver; freshly excised porcine Iiver; high-intensity focused ultrasound exposure; local tissue motion; motion-compensated decorrelation; tissue; tissue coagulation; ultrasonic echoes; Acoustics; Correlation; Decorrelation; Educational institutions; Electronic mail; Lesions; Radio frequency; coagulation; correlation; high-intensity focused ultrasound (HIFU); lesion detection;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Ultrasonics Symposium (IUS), 2012 IEEE International
Conference_Location :
Dresden
ISSN :
1948-5719
Print_ISBN :
978-1-4673-4561-3
Type :
conf
DOI :
10.1109/ULTSYM.2012.0654
Filename :
6562419
Link To Document :
بازگشت