• DocumentCode
    2695019
  • Title

    Stimulation of pulsed-periodical diffuse discharge to be used for medical instrumentation sterilisation

  • Author

    Shcolnikov, E. Ya ; Garkusha, O.V. ; Maslennikov, S.P. ; Netchaev, N.N. ; Novozhilov, A.E. ; Lai, G.Y.

  • Author_Institution
    Moscow Eng. Phys. Inst., Russia
  • Volume
    2
  • fYear
    2003
  • fDate
    15-18 June 2003
  • Firstpage
    1120
  • Abstract
    Gas-discharge methods of the contaminated matter sterilization have some substantial merits. Among them the low sterilization temperature and high efficiency should be mentioned. The most promising method seems to be sterilization with the pulse-periodic diffuse discharge. For the study of the diffuse discharge initiation an experimental equipment has been developed which included gas-discharge chambers with various electrode systems and the high voltage pulse-periodic generator. The common feature of all electrode systems is their ability to form highly nonuniform electric fields. The pulse-periodic source has been developed on the basis of a pulsed transformer and a high voltage block for the formation of nanosecond pulses with the amplitude up to 120 kV and front duration up to 10 ns. The discharge electric power supply system includes a direct voltage bias. The diffuse discharge was exited in the pressure range from 1.3 kPa to 26.7 kPa, its volume was up to 20 dm/sup 3/. The discharge parameters are dependent on the air pressure in the chamber, the discharge gap configuration and the generator pulse amplitude. The essential role of the bias voltage was observed. It contributes to more uniform structure of the discharge in larger spaces.
  • Keywords
    discharges (electric); electric fields; high-voltage techniques; pulse generators; pulse transformers; pulsed power supplies; 1.3 to 26.7 kPa; 10 ns; 120 kV; discharge electric power supply system; discharge gap configuration; electrode systems; gas-discharge methods; generator pulse amplitude; high voltage pulse-periodic generator; medical instrumentation sterilisation; nonuniform electric fields; pulse-periodic diffuse discharge; pulsed transformer; pulsed-periodical diffuse discharge; sterilization temperature; voltage bias; Electrodes; Fault location; Instruments; Nonuniform electric fields; Power supplies; Pulse generation; Pulse transformers; Pulsed power supplies; Temperature; Voltage;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Pulsed Power Conference, 2003. Digest of Technical Papers. PPC-2003. 14th IEEE International
  • Conference_Location
    Dallas, TX, USA
  • Print_ISBN
    0-7803-7915-2
  • Type

    conf

  • DOI
    10.1109/PPC.2003.1278008
  • Filename
    1278008