A phantom study to cross-validate multimodality shear wave elastography techniques
Author :
Hua Xie ; Shamdasani, Vijay ; Heng Zhao ; Pengfei Song ; Shiwei Zhou ; Robert, Jean-Luc ; Greenleaf, James ; Shigao Chen
Author_Institution :
Philips Res. North America, Briarcliff Manor, NY, USA
fYear :
2012
fDate :
7-10 Oct. 2012
Firstpage :
1858
Lastpage :
1861
Abstract :
As various quantitative shear wave elastography techniques are being implemented commercially and used clinically, it is important to understand what elastic modulus is measured by those techniques and to assess technical factors influencing measurement accuracy. Cross-validated in this study were four shear wave elastography techniques including ultrasound elastography point quantification (ElastPQ), magnetic resonance elastography (MRE), 1-D transient elastography (TE), and a Verasonics ultrasound system based elastography feature. ElastPQ is an acoustic radiation force (ARF) based technique implemented on an ultrasound system iU22 for a curvilinear probe C5-1. MRE measurements were made at a shear wave frequency of 100 Hz. The 1-D TE system uses an external mechanical vibrator at 100 Hz to generate shear waves and a 7.5 MHz single element transducer to track shear wave propagation. The Verasonics ultrasound system with a linear probe L7-4 uses a plane wave transmit imaging mode to measure propagation speed of shear waves induced by ARF. Six custom-made elasticity phantoms were measured selectively and were assumed purely elastic by the four techniques. The results show good agreement between two ARF based techniques: ElastPQ and Verasonics based feature. When comparing ElastPQ to 1-D TE, the discrepancy for the hardest phantom could be explained by the diffraction artifact in the latter system. Further improvements on the 1D-TE reconstruction algorithm are ongoing. The overestimation by ElastPQ and Verasonics against MRE could be attributed to frequency dispersion due to slight viscosity in phantoms. MRE measures shear wave speed at a single frequency 100 Hz, whereas the other two estimate an effective group velocity over a larger bandwidth for a broadband shear wave signal. 1-D TE and MRE measurements are in closest agreement due to similar shear wave frequencies. More validation of these techniques is required on real patients in clinical settings.
Keywords :
biomechanics; biomedical ultrasonics; elastic moduli; elastic waves; phantoms; ultrasonic propagation; 1D transient elastography; 1D-TE reconstruction algorithm; ElastPQ; Verasonics ultrasound system based elastography feature; acoustic radiation force based method; clinical settings; cross-validate multimodality shear wave elastography; curvilinear probe C5-1; elastic modulus; external mechanical vibrator; frequency 100 MHz; frequency 7.5 MHz; magnetic resonance elastography; phantom; plane wave transmit imaging mode; quantitative shear wave elastography; shear wave propagation; single element transducer; ultrasound elastography point quantification; Acoustic measurements; Acoustics; Frequency measurement; Phantoms; Ultrasonic imaging; Ultrasonic variables measurement; magnetic resonance elastography; shear modulus; shear viscosity; shear wave dispersion; ultrasound shear wave elastography;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Ultrasonics Symposium (IUS), 2012 IEEE International
Conference_Location :
Dresden
ISSN :
1948-5719
Print_ISBN :
978-1-4673-4561-3
Type :
conf
DOI :
10.1109/ULTSYM.2012.0466
Filename :
6562453
Link To Document :
بازگشت