DocumentCode
2697579
Title
Bipolar electrosurgery with long electrodes for RF coagulation of atrial tissue
Author
Berjano, E.J. ; Hornero, F. ; Oria, Carlos ; Montero, Anastasio
Author_Institution
Lab. Integrado de Bioeng., Valencia Univ., Spain
Volume
6
fYear
1997
fDate
30 Oct-2 Nov 1997
Firstpage
2528
Abstract
We performed experiments on porcine atrial tissue fragments with a custom-built bipolar electrosurgical forceps fitted with long electrodes (20 mm) in order to create long and transmural lesions. Results show that impedance monitoring during heating and visual supervision of the lesion zone, are optimum parameters to control lesion evolution and avoid overheating in the tissue. Transmural lesions without charring were created with 15 watts of power during 10-15 seconds on initial impedances between 50-90 ohm
Keywords
bioelectric phenomena; biological tissues; biomedical electrodes; cardiovascular system; coagulation; surgery; 10 to 15 sec; 15 W; 50 to 90 ohm; RF coagulation; atrial fibrillation; atrial tissue; bipolar electrosurgery; electrosurgical forceps; impedance monitoring during heating; lesion evolution control; long electrodes; long lesions; optimum parameters; porcine tissue; surgical maze; transmural lesions; visual supervision; Catheters; Coagulation; Electrodes; Heart; Laboratories; Lesions; Radio frequency; Surface impedance; Surgery; Temperature control;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 1997. Proceedings of the 19th Annual International Conference of the IEEE
Conference_Location
Chicago, IL
ISSN
1094-687X
Print_ISBN
0-7803-4262-3
Type
conf
DOI
10.1109/IEMBS.1997.756844
Filename
756844
Link To Document