Title
Navigated Control Functional: Combining surgical navigation and real-time nerve activity measurement for a drill´s motor control
Author
Shi, Jiaxi ; Heininger, Sebastian ; Lueth, Tim C.
Author_Institution
Inst. for Micro Technol. & Med. Device Technol. (MIMED), Tech. Univ. of Munich, Garching, Germany
fYear
2011
fDate
9-13 May 2011
Firstpage
1
Lastpage
4
Abstract
This paper presents a new method Navigated Control Functional for protecting a nerve from damage during surgery. This method combines surgical navigation and realtime nerve activity measurement for power control of a surgical drill´s motor. The power of a drill´s motor is automatically reduced when the nerve activity is identified as critical or the drill reaches a predefined prohibited area. The purpose of this method is to increase the safety of patients (especially nerves) and the ergonomics of surgeons and also to shorten the operation time. Furthermore, a laboratory experiment shows the feasibility of this method in principle.
Keywords
biomedical measurement; medical image processing; neurophysiology; power control; real-time systems; stereo image processing; surgery; navigated control functional method; nerve damage protection; patient safety; real-time nerve activity measurement; surgeon ergonomics; surgery; surgical drill motor power control; surgical navigation;
fLanguage
English