• DocumentCode
    2713982
  • Title

    Low-power systematic integration of the instrumentation amplifier and the OTA-C filter for ECG acquisition systems

  • Author

    Tsai, Tsung-Heng ; Chen, Yen-Cheng

  • Author_Institution
    Electr. Eng. Dept., Nat. Chung Cheng Univ., Ming-Hsiung, Taiwan
  • fYear
    2011
  • fDate
    3-5 Nov. 2011
  • Firstpage
    53
  • Lastpage
    56
  • Abstract
    This paper presents low-power systematic integration of the instrumentation amplifier and the low-pass filter for portable ECG monitoring systems. To reduce overall power consumption, the electrode offset cancellation was carried out by sharing one operational transconductance amplifier (OTA) between the amplifier and the filter. Dual loops were implemented to provide offset cancellation so that the cancellation rage is extended. The analog front-end circuit is designed and simulated in a TSMC 90 nm CMOS process. The total power consumption is 3.8 μW from a 1-V supply voltage. Simulation results show that the achieved CMRR is 100.3 dB.
  • Keywords
    CMOS analogue integrated circuits; biomedical electrodes; biomedical electronics; electrocardiography; instrumentation amplifiers; low-pass filters; low-power electronics; medical signal processing; operational amplifiers; patient monitoring; portable instruments; power consumption; CMOS; OTA-C filter; TSMC; analog front-end circuit; common-mode rejection ration; electrode offset cancellation; instrumentation amplifier; low-pass filter; low-power systematic integration; operational transconductance amplifier; portable ECG monitoring systems; power consumption; Electrocardiography; Electrodes; Instruments; Low pass filters; Power demand; Transconductance; Transistors;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Bioelectronics and Bioinformatics (ISBB), 2011 International Symposium on
  • Conference_Location
    Suzhou
  • Print_ISBN
    978-1-4577-0076-7
  • Type

    conf

  • DOI
    10.1109/ISBB.2011.6107643
  • Filename
    6107643