Title :
Pulse Oximeter Signal Amplitudes in Different Body Parts for Wireless Solutions
Author :
Konttila, Antti ; Määttälä, Miia ; Alasaarela, Esko
Author_Institution :
Univ. of Oulu, Oulu
Abstract :
Purpose of this study was to measure pulse oximeter signal amplitudes in different parts of human body. Research results can be utilised in the future when choosing measurement site for wireless continuously monitoring probe. To enable the measurements, pulse oximeter with special characteristics was designed and built, because no suitable instrumentation for this kind of experiments was available. Measurements were made from 20 different locations. The best pulse signals were obtained from thumb and nose. According to this study minimum thresholds can be considered to be for normalized amplitude 0.02 and for SNR 20 dB.
Keywords :
biochemistry; oximetry; patient monitoring; human body; pulse oximeter signal amplitude measurement; wireless monitoring probe; Blood; Instruments; Monitoring; Optical devices; Optical pulses; Optical receivers; Optical transmitters; Probes; Pulse measurements; Senior citizens;
Conference_Titel :
Future Generation Communication and Networking (FGCN 2007)
Conference_Location :
Jeju
Print_ISBN :
0-7695-3048-6
DOI :
10.1109/FGCN.2007.188