• DocumentCode
    2716157
  • Title

    Modeling of a capacitive link for data telemetry to biomedical implants

  • Author

    Takhti, Mohammad ; Asgarian, Farzad ; Sodagar, Amir M.

  • Author_Institution
    ECE Dept., K.N. Toosi Univ. of Technol., Tehran, Iran
  • fYear
    2011
  • fDate
    10-12 Nov. 2011
  • Firstpage
    181
  • Lastpage
    184
  • Abstract
    This paper presents modeling of a capacitive link for wireless data transfer to implantable biomedical microsystems. Based on the proposed model, voltage transfer ratio of the link is calculated, and its most important parameters are highlighted. To validate the suggested model, measurements are performed with 5 mm × 5mm plates, which have a layer of chicken breast with 3.5 mm thickness in between. Comparison between experimental results and model-based simulations shows that the model can be used for characterizing capacitive links.
  • Keywords
    biomedical telemetry; capacitors; data communication; prosthetics; biomedical implants; capacitive link modeling; chicken breast; data telemetry; implantable biomedical microsystems; voltage transfer ratio; wireless data transfer; Biological system modeling; Capacitance; Conductivity; Couplings; Data models; Implants; Skin;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Biomedical Circuits and Systems Conference (BioCAS), 2011 IEEE
  • Conference_Location
    San Diego, CA
  • Print_ISBN
    978-1-4577-1469-6
  • Type

    conf

  • DOI
    10.1109/BioCAS.2011.6107757
  • Filename
    6107757