Title
An Economical Multichannel Integrated Receiver/Reconstruction System for MRI
Author
Dalal, Ishaan L.
Author_Institution
Dept. of Electr. Eng., Cooper Union, New York, NY
fYear
2006
fDate
38899
Firstpage
1
Lastpage
2
Abstract
A design is presented for a multichannel integrated receiver/reconstruction system for MRI that alleviates the concerns associated with commercial solutions regarding cost, scalability and the ability to access intermediate data. Up to 16 MR signals are sampled directly at RF, with digital down-conversion and image reconstruction performed by a single-chip field-programmable gate array (FPGA). Imaging researchers can readily access data from any point in the processing chain over a network, and the design can scale modularly in blocks of 8n or 16n channels
Keywords
biomedical MRI; field programmable gate arrays; image reconstruction; medical image processing; FPGA; MRI; digital down-conversion; economical multichannel integrated receiver/reconstruction system; image reconstruction; single-chip field-programmable gate array; Band pass filters; Coils; Costs; Demodulation; Field programmable gate arrays; Image reconstruction; Magnetic resonance imaging; Operational amplifiers; RF signals; Radio frequency;
fLanguage
English
Publisher
ieee
Conference_Titel
Life Science Systems and Applications Workshop, 2006. IEEE/NLM